OPERATION
EDIT WHILE A PROGRAM IS RUNNING
(Figure .1)
14
1 . . .To .identify .which .program .is .running, .press .PROGRAM .NUMBER .
NOTE: The program number cannot be changed using while the program is running.
2 . . .To .determine .the .time .remaining .for .the .completion .of .the .program, .press .NIGHT .TIME .(DELAY .START) .
. . . . .The .time .remaining .appears .on .the .Main .Display .for .5 .seconds .
3 . . .Any .individual .parameter .can .be .increased .or .decreased .during .the .actual .running .of .all .30
. . . . .programs .
NOTE: HEAT RATE cannot be changed to "NO" in the stage currently running or in the
stages already completed.
4 . . .To .change .a .parameter .while .a .program .is .running, .press .ENTER ./ .REVIEW .to .advance .to .the .desired . .
. . . . .STAGE .and .parameter .(i .e . .HEAT .RATE, .TEMP .or .HOLD .TIME) . . .Initially, .STAGE .1 .will .appear . . . . .
. . .
. . . . .Any .parameter .can .be .increased .or .decreased .by .pressing .
or .
.
5 . . .Any .program .can .be .stopped .or .started .by .pressing .the .START ./ .STOP .button .
6 . . .If .a .program .is .edited .while .running .and .the .HEAT .RATE .in .STAGE .2, .3 .or .4 .is .set .to ."COOL" .but .the .
. . . .corresponding .TEMP .programmed .is .entered .to .heat .to .a .higher .temperature .than .the .previous .stage, . . .
. . . .the .Infinity .will .try .to .heat .with .a .0
°
F ./ .Min .(0
°
C ./ .Min) .HEAT .RATE . . .In .this .case, .when .NIGHT . . .
. . . .TIME .(DELAY .START) .is .pressed, .99:99 .(Hr .: .Min) .flashes .on .the .Main .Display, .indicating .that .the . .
. . . .program .cannot .be .completed .