To .start .immediately, .press .START ./ .STOP .
Follow .the .same .programming .procedures .if .a .three .or .four .stage .program .is .desired .
Press .NIGHT .TIME .(DELAY .START) .to .start .an .overnight .burnout . . .(See .SAMPLE .ON .STAGE .PROGRAM .to .pro-
gram .NIGHT .TIME .(DELAY .START) .time .
NOTE:
When the temperature is set to a lower temperature than in the previous stage,
the furnace ignores the programmed HEAT RATE (except for "NO") and
automatically cools to the pre-set temperature. The time required for the cooling
stage is determined by the pre-set temperature. The lower the temperature,
the more time needed to cool down.
OPERATION
HEAT RATE
TEMP
HOLD TIME
STAGE 1
STAGE 2
STAGE 3
STAGE 4
10
°
F .(6
°
C)
1600
°
F .(871
°
C)
1:00
NO
NO
*
*
*
*
* .Though .any .number .may .appear, .the .unit .is .deactivated .when .“NO” .is .selected .for .the .HEAT .RATE .
20
°
F .(11
°
C)
600
°
F .(316
°
C)
30
16
SAMPLE THREE STAGE PROGRAM
(Program .with .Cooling) .