background image

10

2. Standard solution of 12.90 mS/cm at 25 °C : Accurately

weight out 7.4365 grams of research grade dried Potassium

Chloride (KCL). Dissolve in 1000 ml of distilled water.

3. Standard solution of 111.9 mS/cm at 25 °C : Accurately

weight out 74.264 grams of research grade dried Potassium

Chloride (KCL). Dissolve in 1000 ml of distilled water.

[Note:

You can store the unused portion of the standard

solution in a plastic container for six months but the air space

between the cap and the solution must be kept to an absolute

minimum. Storing the excess solution below 4 °C can

increase the storage life. If you have any doubt of the

accuracy of the stored solution, a fresh batch should be

prepared.

]

Summary of Contents for 3011M

Page 1: ...Operation Manual MODEL 3011M Basis Portable Cond TDS Salt Temperature Meter 3011M...

Page 2: ...SPLAY KEYS FUNCTIONS 4 A Display 4 B Operational keys description 5 OPERATIONAL PROCEDURES 6 A Calibration 6 B Conductivity Salinity TDS Measurements 8 C Save Recall and Delete Data 8 D Preparing Cond...

Page 3: ...th user prompts and mode indicators The unit prompts the user through calibration and measurement procedures The unit is also equipped with a non volatile memory allowing the user to store 50 differen...

Page 4: ...he procedure outlined below Figure 1 Battery compartment 1 Use a screw driver to remove the four screws and battery cover to expose the battery compartment Figure 1 2 Replace the 9 V battery 3 Replace...

Page 5: ...ode 5 Main display for Conductivity Salinity or TDS values 6 mg L Milligrams Liter indicates TDS measurement 7 Indicates Temperature Coefficient 8 Reference temperature unit 9 Data storage site number...

Page 6: ...of data from the data storage 18 ALL To delete all the data in the data storage 19 CELL K Conductivity cell constant value 20 FULL This will indicate that all 50 data storage sites are used up 21 ppt...

Page 7: ...elete Each and Delete All mode In the Delete Each mode view the to be deleted data and data site numbers by pressing these keys 4 In measure mode press this key to go to the Calibration mode 5 In the...

Page 8: ...00 Press Enter key to save the new value and go to the next calibration parameter Probe Basic Cell Constant The main display shows the Cell Constant of the conductivity probe calibrated previously or...

Page 9: ...Press the Enter key to save and exit the calibration B Conductivity Salinity TDS Measurements 1 Power up the unit Connect the conductivity probe to the unit 2 Press Mode key to go to Uncompensated co...

Page 10: ...go into the Delete mode b Select Delete All or Delete Each mode by pressing the Up or Down key c In the Delete All mode press Enter key to clear all stored data Deletion is now complete d In the Delet...

Page 11: ...rams of research grade dried Potassium Chloride KCL Dissolve in 1000 ml of distilled water Note You can store the unused portion of the standard solution in a plastic container for six months but the...

Page 12: ...alue input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 10 0 to 120 0 C 1 Incorrect standard solution 2 Conductivity cell contam...

Page 13: ...9 mg L 4 75 47 49 g L 47 5 200 0 g L 0 1 mg L 1 mg L 0 01 g L 0 1 g L 0 5 F S TDS K 0 1 0 000 9 999 mg L 10 00 99 99 mg L 100 0 400 0 mg L 0 001mg L 0 01mg L 0 1 mg L TDS K 1 0 0 0 999 9 mg L 1000 999...

Page 14: ...warranty please contact them to notify us of the situation Jenco Service Department alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be...

Page 15: ...om Jenco Electronics Ltd 6F NO 81 Sec 2 Chang an E Rd Jhongshan District Taipei City 104 Taiwan TEL 886 2 2508 2928 FAX 886 2 2508 2938 E Mail sales jenco com tw Website www jenco com tw Shanghai Jenc...

Reviews: