background image

7

access these sections:

1. Power up the unit. Connect the conductivity probe to the unit.

2. Press

“Mode”

key to go to conductivity mode. “ATC” icon,

“uS/cm” icon and “AUTO” icon will lit up.

3. Rinse the conductivity probe in distilled water and immerse

them in the standard solution. Allow temperature reading to

stabilize, then press

“Cal”

key to enter the calibration mode.

“CAL” and “TDS” icons appears on the LCD.

[

Note

: Press

“Enter”

key to accept any values changes in

each section and automatically advance to the next section. If

there are no changes, the unit accepts the current value and

proceeds to the next section.]

TDS Constant

TDS is determined by multiplying conductivity (mS/cm) by a

TDS factor. The default factor value is 0.65. To change the

TDS factor, use the

“Up”

or

“Down”

keys to adjust the value

between 0.30 to 1.00. Press

“Enter”

key to save the new

value and go to the next calibration parameter.

Probe Basic Cell Constant

The main display shows the Cell Constant of the conductivity

probe (calibrated previously or default, the deviation range is

70% to 130%). The secondary display shows the current

selected cell constant which is either 0.1, 1.0, 10.0 or 475.

Press the

“Up”

or

“Down”

keys to select the one you want to

use. Press

“Enter”

key to save the new value and go to the

next calibration parameter.

Temperature Coefficient

The unit uses the temperature coefficient to calculate

temperature compensated conductivity. The default value is

1.91 %/°C. To change the Temperature Coefficient, use the

“Up”

or

“Down”

keys to adjust the value between 0.00 to

4.00 %/°C. Press

“Enter”

key to save the new value and go

to the next calibration parameter.

Temperature Reference

The unit uses the temperature reference value to calculate

temperature compensated conductivity. The default value is

25 °C. To change the temperature reference, press the

“Up”

Summary of Contents for 3011M

Page 1: ...Operation Manual MODEL 3011M Basis Portable Cond TDS Salt Temperature Meter 3011M...

Page 2: ...SPLAY KEYS FUNCTIONS 4 A Display 4 B Operational keys description 5 OPERATIONAL PROCEDURES 6 A Calibration 6 B Conductivity Salinity TDS Measurements 8 C Save Recall and Delete Data 8 D Preparing Cond...

Page 3: ...th user prompts and mode indicators The unit prompts the user through calibration and measurement procedures The unit is also equipped with a non volatile memory allowing the user to store 50 differen...

Page 4: ...he procedure outlined below Figure 1 Battery compartment 1 Use a screw driver to remove the four screws and battery cover to expose the battery compartment Figure 1 2 Replace the 9 V battery 3 Replace...

Page 5: ...ode 5 Main display for Conductivity Salinity or TDS values 6 mg L Milligrams Liter indicates TDS measurement 7 Indicates Temperature Coefficient 8 Reference temperature unit 9 Data storage site number...

Page 6: ...of data from the data storage 18 ALL To delete all the data in the data storage 19 CELL K Conductivity cell constant value 20 FULL This will indicate that all 50 data storage sites are used up 21 ppt...

Page 7: ...elete Each and Delete All mode In the Delete Each mode view the to be deleted data and data site numbers by pressing these keys 4 In measure mode press this key to go to the Calibration mode 5 In the...

Page 8: ...00 Press Enter key to save the new value and go to the next calibration parameter Probe Basic Cell Constant The main display shows the Cell Constant of the conductivity probe calibrated previously or...

Page 9: ...Press the Enter key to save and exit the calibration B Conductivity Salinity TDS Measurements 1 Power up the unit Connect the conductivity probe to the unit 2 Press Mode key to go to Uncompensated co...

Page 10: ...go into the Delete mode b Select Delete All or Delete Each mode by pressing the Up or Down key c In the Delete All mode press Enter key to clear all stored data Deletion is now complete d In the Delet...

Page 11: ...rams of research grade dried Potassium Chloride KCL Dissolve in 1000 ml of distilled water Note You can store the unused portion of the standard solution in a plastic container for six months but the...

Page 12: ...alue input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 10 0 to 120 0 C 1 Incorrect standard solution 2 Conductivity cell contam...

Page 13: ...9 mg L 4 75 47 49 g L 47 5 200 0 g L 0 1 mg L 1 mg L 0 01 g L 0 1 g L 0 5 F S TDS K 0 1 0 000 9 999 mg L 10 00 99 99 mg L 100 0 400 0 mg L 0 001mg L 0 01mg L 0 1 mg L TDS K 1 0 0 0 999 9 mg L 1000 999...

Page 14: ...warranty please contact them to notify us of the situation Jenco Service Department alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be...

Page 15: ...om Jenco Electronics Ltd 6F NO 81 Sec 2 Chang an E Rd Jhongshan District Taipei City 104 Taiwan TEL 886 2 2508 2928 FAX 886 2 2508 2938 E Mail sales jenco com tw Website www jenco com tw Shanghai Jenc...

Reviews: