background image

innoCon 6800 Monitor Manual

The instrument is compliant with standard Modbus-

RTU, with a fixed serial 

32767), 

The host computer sends orders in the following format:

This is the first data being ac

cessed. Please refer to the following table.

Instrument ID

address

Instrument ID

address

Length

Example

Length

Example

0x01

0x03

0x0001

0x0001

0xD5CA

1 byte

1 byte

2 byte

2 byte

2 byte

Order

Order

of data

data

data

Data content

CRC16

Page 17

 :

1

(

RS485

)

The client computer responds in the following format:

When the instrument displays

 02, it means the address is incorrect.

0x01

0x03

0x02

0x02

0xBC

0xB895

1 byte

1 byte

1 byte

N byte

2 byte

CRC16

 :

1

Data addresses:

 
  
 
 
 
 
 
 
 
 
 
 

(01)

 

0x01

  DO measurement value

(02) 

0x02

  DO current  

(03) 

0x03

  Temperature value

(04)  

0x04

  Temperature current

Reading value

 x 0.1

x 0.01

Reading value

 x 0.01

Reading value

 x 0.1

Reading value

 x 0.01

www.jensprima.com

jens

Summary of Contents for innoCon 6800O

Page 1: ...WALL MOUNTING JENSPRIMA INSTRUMENTS Online Dissolved Oxygen Analyzer Operating Instruction innoCon 6800O PRIMA_O2_3BC6051 NORMAL ATC REL WASH 4 88mg l C 25 0 1 2 O O2...

Page 2: ......

Page 3: ...r Manual www jensprima com Contents Safety 2 3 5 5 6 7 8 9 10 11 12 13 13 16 17 Display Maintenance General 6 5 Analysis System 14 14 14 15 Zero Calibration Concentration Calibration Output Interface...

Page 4: ...factured by jensprima Damages to the instrument caused by the use of sensors not manufactured by jensprima within the warranty period will not be covered by the warranty Only technicians authorized by...

Page 5: ...ion methods standard 4 20mA RS485 output and a variety of control function output Both Chinese and English menus users can do programming on their own Sewage andindustrial wastewater treatment sh farm...

Page 6: ...r Temperature load 600 RS485 Modbus RTU IP65 138 x 138mm 144 x 144 x 110mm innoSens 420 DO Electrode Suitable for measuring dissolved oxygen in 0 20 00mg L ppm water such as sewage treatment aquacultu...

Page 7: ...138mm on the instrument case put the instrument into the case from the front of the panel of the case insert screws from the back cover and then lock the fastening block to the fastening bar from the...

Page 8: ...25 Note the electrodes innoCon 6800 Monitor Manual Page 6 PRIMA_pH_1BC6051 NORMAL ATC REL WASH 7 00 pH C 25 0 1 2 O innoSens 425 0 20 00ppm 0 200 0ppb Immersion type O ring O ring Molded shell electro...

Page 9: ...ith special training Notes Power supply must be cut o otherwise the instrument and the sensor are likely to be damaged Page 7 power 90 260Vac sensor live wire of power supply neutral wire of power sup...

Page 10: ...atus of the washing relay ATC MTC manual Menu key Back key Enter key Arrows key Return to previous menu Accept the gure input or menu chosen or o the backlight of the display Entry into the setup mode...

Page 11: ...2 4 00mA 1 AnalysisSystem Temperature System Measurement System Pressure Salinity Damping 1 Output Analog current 1 Analog current 2 Relay1 Relay2 Relay3 1 General Language Serial Number Restore 1 Int...

Page 12: ...previous menu 2 Select Restore and press Enter 3 Press upward and downward arrows to select YES or NO and press Enter General 2 Select Language press Enter and you will get a list of available languag...

Page 13: ...Enter to save the setting and then reture to previous menu Measurement System Analysis System Page 11 Temperature System Temp Unit Pt 1000 0 0 025 0 Sensor Type Temp Offset Manual Temp 1 Measurement...

Page 14: ...n reture to previous menu Analog Current2 4 00mA 20 00mA 100 0 0 00mA 000 0 Optional range 10 0 110 0 14 0 230 0 Optional range 0 0 120 0 32 0 248 Optional range 1 00 1 00mA mA Offset 1 Relay 1 1 Ente...

Page 15: ...y 1 Relay 2 Press Enter to test whether Relay3 is closed or open Press Enter to output 4mA 12mA 20mA compulsorily Press Enter to output 4mA 12mA 20mA compulsorily Press Enter to test whether Relay2 is...

Page 16: ...tore 1 Zero Calibration Current Current nA value of the electrode Enter the current standard temperature setting range available 0 60 0 C Current nA value of the electrode Current temperature of stand...

Page 17: ...nd Saturation value and Press ENTER 80 0nA 25 0 100 1 1 Wash the electrode dry and put it into saturated air input the current saturated air temperature the next interface Saturation Calibration Curre...

Page 18: ...is used after a few months storage replace the electrolyte rst before using If the membrane cap fails eg the response time becomes longer the value deviationbecomese larger or the calibration does not...

Page 19: ...0xD5CA 1 byte 1 byte 2 byte 2 byte 2 byte Order Order of data data data Data content CRC16 Page 17 1 RS485 The client computer responds in the following format When the instrument displays 02 it means...

Page 20: ...Jensprima Instruments Shanghai Co Ltd Web www jensprima com Email info jensprima com 2017 Jensprima all rights reserved Informa on contained in this document is subject to change without no ce...

Reviews: