innoCon 6800 Monitor Manual
Output
Analog current 1 ((for pH/ORP current output))
2. Select “Analog curent1” and press “Enter”.
previous menu.
Page 12
Analog Current1
4.00mA
:
20.00mA :
+14.00
+0.00mA
+00.00pH
pH
Optional ranges::-2.00~15.00pH、-2000~1900mV
Optional ranges::-1.00~16.00pH、-1900~2000mV
Optional ranges::
-1.00~1.00mA
mA Offse:
1
▲
▲
Analog current 2 (for temperature current output)
2. Select “Analog curent2” and press “Enter”.
previous menu.
Analog Current2
4.00mA
:
20.00mA :
+100.0℃
+0.00mA
+000.0℃
Optional ranges::-10.0~110.0℃、14.0~230.0℉
Optional ranges::0.0~120.0℃、32.0~248℉
Optional ranges::-1.00~1.00mA
mA Offset:
1
▲
▲
Relay 1
2. Select “Relay1” and press “Enter”.
previous menu.
Relay 1
ON/OFF
:
Action :
HI
+10.00pH
0.10pH
ON
Options:ON/OF
Options:High /Low
Optional ranges:-2.00~16.00pH、-2000~2000mV
Optional ranges:0~2.00pH、0~200mV
Set point
:
Hysteresis :
1
▲
▲
Relay 2
2. Select “Relay2” and press “Enter”.
previous menu.
Relay 2
ON/OFF
:
Action :
LO
+4.00pH
0.10pH
ON
Options: ON/OFF
Options: HI/LO
Optional ranges:2.00~16.00pH、-2000~2000mV
Optional ranges:0~2.00pH、0~200mV
Set point
:
Hysteresis:
1
▲
▲