Set the Signal to 

SEI 

, if not already  

(You can change it to 

BEIW

, if desired later). 

16.

 

Click the 

HT Ready

 

icon to turn on the High Voltage, now should be 

  

Some intensity should be seen in the image window.  If not press 

View

 

 and 

ACB

.  

ACB

 takes several seconds to run. Something should be visible on the 

screen  now. 

17.

 

Click the 

Gun

 icon 

.  The electron gun and alignment dialog appears.  

 
There are a couple of options at this point depending on skill level, needs, etc.  
Basically you need to check gun alignment and then filament saturation. 
 

Recommended Procedure:

 Manually perform this task. It only takes a minute and 

will insure optimized operation and best images. Note: We do NOT use Auto filament 
heating or Auto Gun Control. We found the use of these seem to shorten filament life 
significantly. 

 

 

a.

 

Select an area on the sample or sample holder that has reasonably 
homogeneous brightness at moderate magnification (1-5 kx) 

b.

 

Adjust Tilt X and Y for maximum average image brightness at a low Spot 
Size setting (<=20) by iteration. 

c.

 

Adjust Shift X and Y for maximum average image brightness at a high Spot 
Size setting (=>60). 

ACB

 or manual adjustment of brightness/contrast will 

likely be needed as the image brightness will usually saturate. If intensity 
disappears before reaching Spot Size 60, adjust shifts at highest spot size 
you still have an image, then increase to above 60 and adjust again. Again 
iterate controls to optimize. 

d.

 

Steps b and c can be iterated to optimize alignment. This is usually not 
necessary unless large adjustments in the first pass were required. 

Summary of Contents for 6060LV

Page 1: ...computer use the account USER jeol PASSWORD jeol Please do not log into other accounts Use portable media ftp or UIUC NetFiles to transfer your data 3 Prepare your samples before venting Please keep...

Page 2: ...the height gauge for desired working distance No part of the sample should extend above the ruler end A 10 mm W D should be used for nearly all work except low magnification imaging or if you need to...

Page 3: ...contact these when observing your sample 10 Turn on the IR chamber scope if off using the switch button on the control box and a switch on CRT monitor and adjust illumination as needed Always use cham...

Page 4: ...filament heating or Auto Gun Control We found the use of these seem to shorten filament life significantly a Select an area on the sample or sample holder that has reasonably homogeneous brightness a...

Page 5: ...ases after going through a maximum before beginning of orange indicator the gun tilt shifts are misaligned This is usually because the filament was significantly under saturated when you started Resta...

Page 6: ...cus 21 Click ACB auto contrast brightness 22 Focus and check the objective aperture centering especially for high magnification work 23 Fine focus and correct the astigmatism the image manually or cli...

Page 7: ...rging artifacts h Many commonly used functions are available on the knob panel magnification brightness contrast stigmator x and y image fine shift x and y scanning mode and auto functions focus stigm...

Page 8: ...chassis stops blinking slide open stage drawer 6 If changing samples go to step 7 of general operation 7 Change vacuum mode to HV if in LV mode 8 Carefully slide the stage drawer closed Press EVAC bu...

Page 9: ...other translation control Iterate to fine tune Correct any severe astigmatism see below to be able to better fine tune 5 Deselect Wobb Manually Correcting the Astigmatism Note You need an image with...

Page 10: ...10 30 kV is the condition the instrument provides it s best resolution 100 kX possible This is a fairly noisy image even on a Au C sample You must choose the spot size according to your sample and wh...

Page 11: ...focus astigmatism correction a little more difficult to do ACB and manual knob adjustment function in all the BEIW modes Selecting the Operating Voltage 1 Click Acc Volt to get 2 Double Click value t...

Page 12: ...samples but this requires addition procedures wet In general use as low as pressure as possible that works for imaging the sample 1 to 10 s of Pa Again BEIW Shadow mode provides secondary electron lik...

Page 13: ...computer is responding use the software restart START Menu If the computer is locked press and hold the power button on the front of the computer until it turns off Release the button then press it a...

Page 14: ...d Manual control buttons Manual adjusting buttons spotsize contrast brightness etc and image size switching button are arranged Image display area A SEM image of 640 480 pixels can be displayed and th...

Reviews: