background image

J E O L   J S M - 6 0 6 0 L V   S C A N N I N G   E L E C T R O N   M I C R O S C O P E  

 

 
 
 
 

 

 

Background

 

 

The scanning electron microscope (SEM) is one of the most versatile instruments for the 
examination and analysis of the microstructural characteristics of solids.  Although the SEM and 
optical microscope share the same primary function – making microstructural features and objects 
visible to the human eye – the scanning electron microscope offers some distinct advantages over 
the optical microscope.   The SEM uses electrons rather than visible light waves (200 – 750 nm 
wavelength) for imaging, which allows for observation of relatively large sample features at low 
magnifications or very fine details (high resolution) at high magnifications.  The SEM also offers a 
large depth of field that provides good focus over rough specimen surfaces.  The large depth of 
focus provides a three-dimensional appearance of the specimen in a SEM compared to the nearly 
planar or two-dimensional imaging found in optical microscopes.  In addition, many attachments 
are available for scanning electron microscopes, including x-ray spectrometers for chemical 
composition analysis, backscattered electron detectors for atomic number contrast, transmitted 
electron detectors, hot and cold stages for microscopic observation of high or low temperature 
phenomena, tensile testing stages for observation of deformation and fracture, and special stages 
for analysis of semiconductor devices.   
 
Disadvantages of the 
scanning electron 
microscope include 
relatively high initial, 
operational, and 
maintenance costs, a high 
vacuum operating 
atmosphere that is 
unsuitable for some 
specimens, and difficulty in 
preparing certain types of 
specimens.  Figure 1 
schematically illustrates 
image formation in the 
optical microscope and the 
scanning electron 
microscope. 

 

 

 

 

 

Figure 1.  Basic image formation in an optical microscope and a 

scanning electron microscope.

Summary of Contents for JSM-6060LV

Page 1: ...JEOL JSM 6060LV SCANNING ELECTRON MICROSCOPE Insert Nickname Here Operating Instructions...

Page 2: ...nformation 4 References 4 3 S A M P L E S Sample Holders 7 Sample Preparation 7 4 O P E R A T I O N Instrument Startup 8 Sample Loading 8 Getting an Image 9 Moving Around 9 Image Scanning 10 Zooming I...

Page 3: ...d a NORAN System 6 elemental analysis system see separate operating instructions for the NORAN system Best of all the JEOL JSM 6060LV scanning electron microscope is user friendly and easy to operate...

Page 4: ...provides a three dimensional appearance of the specimen in a SEM compared to the nearly planar or two dimensional imaging found in optical microscopes In addition many attachments are available for sc...

Page 5: ...umn in an electron beam they are controlled and directed to the specimen by a series of electromagnetic lenses and apertures When the electrons in the electron beam hit the specimen a number of electr...

Page 6: ...arrangements may be made talk to Professor Stolk Other types of sample holders are available See Professor Stolk for information on ordering sample holders Sample Preparation Conductive Materials 1 S...

Page 7: ...er 4 After coating your sample use conductive carbon tape carbon tabs carbon paint or copper tape to attach your sample to a sample holder or cylindrical sample mount If you use carbon paint please wa...

Page 8: ...mber is evacuated EVAC button on the front of the instrument is lit green press and hold the VENT button until it lights up yellow This will vent the chamber 2 Wait for the VENT button to stop flashin...

Page 9: ...COARSE button and adjusting the focus knob on the keypad 6 Click on the blue ACB to automatically adjust the contrast and brightness 7 Click on the blue AS Stigma button for automatic astigmatism corr...

Page 10: ...scan rate buttons to see what they do Scan 2 is a TV like mode Scan 3 is a slower scan with higher resolution Zooming In or Out To change magnification use the MAGNIFICATION knob on the keypad or the...

Page 11: ...bar buttons Scan3 and Scan4 are the only modes recommended for photography Images taken on the Scan4 mode are less grainy than those acquired in Scan3 mode The Scan Speed is adjusted in the Scan Scan...

Page 12: ...lly adjust the focus and stigmation settings at a higher magnification It is good practice to focus the image at a higher magnification then back down to the desired photo magnification 3 Adjust the c...

Reviews: