The Advanced tab window is used to set parameters in detail. All o
window must be specified.
■
Checking before an approach
Before performing an approach, check whether the preparation has been completed
according to the procedure in Sect. 5.3, “Preparing for Measurement”.
•
Is the cantilever for the AC mode properly mounted on the cantilever holder?
•
Has coarse approach been executed in such a way that the distance between the
cantilever tip and the specimen surface is 0.3 mm or less?
•
Is the Z stage position in the approach-possible range?
•
Is the AFM value of the AFM amplifier unit about 0 V?
•
Is the FFM value of the AFM amplifier unit about 0 V?
After checking these points, set the indicator selection knob on the AFM amplifier
unit to “RMS”.