FFM observation
5-86
TMPM5200-2
I
nning direction is parallel to the
canti
rce
M
irection is orthogonal to the
canti
sted during
scann
canti
n the ordinary contact-mode AFM, the specimen-sca
lever axis as shown in the left figure below. In the case of FFM (Friction Fo
icroscope), on the other hand, the specimen-scanning d
lever axis as shown in the right figure below. The cantilever is twi
ing due to the friction force occurring between the specimen surface and the
lever tip. The FFM measures the friction force from this torsion of the cantilever.
Scanning direction
Cantilever
Ordinary contact mode AFM
FFM
The following is the procedure for FFM observation.
1.
Select the FFM radio button at the left bottom of the Contact Mode tab
window.
ts, and a topography image and an FFM image are displayed on the
Display Window.
2.
Click on the Scan button.
Scanning star
Topography
image
FFM image
The scanning direction automatically changes by 90
°
when FFM is selected.
The white locations in the FFM image indicate the areas where the friction force is
large.