MEASUREMENT OPERATION
TMPM5200-2
5-257
OSC
Set t
by cl
Output
Set t
outp
If you select ACosT, ASinT, the detection signals in the lock-in amplifier are decomposed
and output into the cos component and sin component, respectively.
If you select Amplitude, Phase, the amplitude
lock-
Auto
W
for t
the i
itself
Advanced
This displays a dialog that enables you to perform a more detailed setting for the lock-in
amplifier.
Level / V
he signal amplitude that is output from the oscillator in V order. You can also set this
icking on the Set button using the slider bar.
he kinds of output signals from the lock-in amplifier. Since the lock-in amplifier can
ut two kinds of output signals, set two signals in pairs.
and phase of the detection signals in the
amplifier are output.
in
igure
Conf
hen you click on this button, the lock-in amplifier sets an appropriate input sensitivity
he current input signal and reference phase so that the reference phase agrees with
nput signal phase by using the automatic adjusting function that the lock-in amplifier
has.
•
Filter
This sets a filter for the input signal. Usually, select Band Pass.
Flat
Do not apply a filter.
Low Pass Apply a low pass filter.
Band Pass Apply a band pass filter.
•
Filter Frequency [kHz]
In the Filter selection, this sets the cutoff frequency if you select Low Pass or Band
Pass.