Setting Channel A
This software oscilloscope enables you to acquire and display data from two channels at
the same time. Channel A is displayed in red, and Channel B in blue. Normally, set
Channel A to Topography (motion in the Z direction of the scanner). The setting of
Channel B differs depending upon the scanning mode (Contact mode, AC Mode, STM,
etc.), so refer to the respective operation methods.
The procedure for setting Channel A is as follows:
1.
Click on the Source
button, and select Topography from the pull-down
menu.
2.
Click on the Coupling
button, and select DC from the pull-down menu.
3.
Click on the Sensitivity
button, and select sensitivity from the pull-down
menu.
The numerical value of the sensitivity displayed in the pull-down menu
automatically changes according to the sensitivity of Z of the presently selected
scanner.
4.
Double-click on the Offset numerical value input box, and enter 0.0 using the
keyboard.
5.
Once you have completed settings, select File from the main menu, and then
select Store Configuration from the pull-down menu to save the set values.
If you skip Step 5 above, the system will revert to the initial settings the next
time it is started.
5.2.5 Lock-In
Amplifier
This optional lock-in amplifier expands the application of the system, enabling you to
perform the following measurements:
•
Viscoeleasticity measurement
•
Lateral modulation friction force measurement
•
SKPM (Scanning Kelvin Probe Microscopy) measurement
•
Electrostatic force measurement
The Model 7265 lock-in amplifier is needed when you perform electrostatic force
measurement.
■
Model 5110 lock-in amplifier
Front operation panel
②
③
④
⑤
⑯
⑫
⑧ ⑩ ⑨
⑪
⑬
⑰ ⑱
⑲ ⑳
①
⑥ ⑦
⑭
⑮
The functions of the switches and indications are detailed below.
TMPM5200-2
5-13