5-36
TMPM5200-2
5.4.2 M
FM (Magnetic Force Microscopy) measurements in AC AMF mode can be performed
ed by clicking on the MFM tab in the SPM
optional MFM cantilever is required. The
and the line-by-line lifting function (lifting at every line) is optionally available.
FM
Tab
M
using the MFM tab window that is open
Parameters window. To do this, however, an
point-by-point lifting function (lifting at every picture element) is included as standard
e observation
■
MFM imag
In general, MFM observations are performed in AC AFM mode. To differentiate the
topographical image of the sample from the MFM image, the topographical image is first
observed at a point and then the cantilever is lifted for observation of the MFM image
formed by the difference of the cantilever phase. The procedure for observing an MFM
image is given below.