TLS5 User Manual
TLS5-UM-00001 Rev 002
20
2.
Apply optical grade isopropyl alcohol or optical grade ethanol to a
small area of a lint-free towel and rub the end of the ferrule over
the wet area.
3.
Wipe the ferrule on a dry area of the lint-free towel.
4.
Using the dusting gas or compressed air, blow the end of the ferrule.
5.
Apply the alcohol or ethanol to a lint-free pipe cleaner or swab and
wipe off the remaining parts of the connector.
6.
With the other end of the pipe cleaner or swab, dry the areas
cleaned.
7.
Using the dusting gas or compressed air, blow the areas cleaned.
Figure 6: Connector (Connector Type May Vary)
Troubleshooting
If any problem described in this section persists, contact JGR Optics or your local
representative.
Connector Issues
Front Panel Connectors
Follow the maintenance procedure described in the section:
“
CLEANING THE
CONNECTOR ENDS”
section on page 18, to ensure that the internal pigtail
connectors are clean and properly connected to the front panel mating sleeve.
If cleaning is not sufficient, the FC/APC connector can be polished.
Other potential Issues
There are very few things that can go wrong with the TLS Tunable Source that
could be solved by the user. Refer to Table 9 on next page for a list of potential
problems and solutions. If problem persist, please contact JGR Optics for
support.
Ferrule
Summary of Contents for TLS5
Page 2: ......