Bit 6, as the service request bit, is set to 1 if a service request has been
generated.
Bit 6, as the master summary bit, is set when there is at least one reason for
the laser source to request service from the controller. That is, the master
summary bit is set if any summary bit in the status byte register is set and if
the corresponding bit in the service request enable register is also set.
•
Bit 7 (operation summary bit) is not used.
Service Request Enable Register
The service request enable register determines which summary bits in the
status byte register can generate service requests. If a summary bit in the status
register is set to 1 and the corresponding bit in the service request enable
register is set to 1, a service request is generated by the laser source. A new
service request is not generated for this condition unless the bit in the status
register or the bit in the service request enable register is cleared and the
condition reoccurs.
Standard Event Status Register
•
Bit 0 (operation complete) is set in response to the
*OPC
common
command. This bit is set when all operations are complete.
Service Request Enable Register
Read with
*SRE?
common query (the value of bit 6 is always 0)
Written to with
*SRE
common command (the value of bit 6 is always
zero, regardless of the value sent with the command)
Cleared by
*SRE
common command with a parameter value of 0
Power-on
Standard Event Status Register
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
power on
(PON)
user
request
(URQ)
command
error
(CME)
execution
error
(EXE)
device
dependent
error
(DDE)
query
error
(QYE)
request
control
(RQC)
operation
complete
(OPC)
Read with
*ESR?
common query
Written to with
Cannot be written to
Cleared by
*ESR?
common query
*CLS
common command
Power-on