PAGE 98 of 177
6.
Move the probe holder arms
and insert the pivot buttons
into the probes pivot button
holes while maintaining the
probes alignment relative to
the yoke’s pivot
(Fig. 230)
.
7.
Tighten the two probe
holder arm adjustment knobs
(Fig. 230)
.
5.8.5. Skew Encoder Cable
The encoder cable of the encoded skew vertical probe holder provides encoded
feedback of the probe holder’s skew angle. Cable routing is at the discretion of
the user. Cable clips have been provided to assist with cable management.
1.
Route the skew encoder cable through any required cable clips
.
2.
Plug the skew encoder cable into
ENC 2
at the rear of the low profile link.
5.8.6. Encoded Skew Vertical Probe Holder Adjustment
To lower the probe
(and probe holder)
to the scan surface, follow these steps:
TIP:
The skew encoder cable removed for illustration purposes.
1.
Lift the probe holder slightly to allow the release of the latch pin
(Fig. 246)
.
Fig. 230 - Clamp probe with arms and tighten knobs
Fig. 231 - Lift probe holder
Fig. 232 - Pull latch pin