PAGE 100 of 177
length of the stroke without latching.
7.
Rotate the latch pin to return to the probe holder to the locking capable
position.
5.8.7. Skew Angle Adjustment
Rotation of the probe holder is possible through adjustment of the skew angle.
1.
Loosen the ratchet lever above the yoke
(Fig. 236)
.
2.
Rotate the yoke
(Up to 90° in either direction)
to the angle required
(Fig. 237)
.
3.
Tighten the ratchet lever
to lock the yoke in place.
Should the ratchet lever
be unable to fully tighten
or release the yoke
(see
Ratchet Lever on page
101 for more information).
4.
The engraved arrow
above the yoke may be
used to align the yoke to
the required degree.
Fig. 236 - Loosen ratchet lever
Fig. 237 - Adjust skew angle
Fig. 238 - Engraved arrow aligns with various degree measurements