PAGE 107 of 177
CX0344 Rev 05.3
1.
Attach the wedges to the probe holders that will be used
(see “Probe Holder
Setup” on page 69 for additional details)
.
2.
Affix the probe holders
(with attached wedges)
to the probe holder frame.
Place the secondary probe holder at the front of the
1
frame and place the
primary probe holders at the rear of the
2
frame bar
(Fig. 254)
.
TIP:
Due to their larger size, scan results are generally improved when
pulling or dragging phased array wedges.
3.
Mount the probe holder frame to the crawler
(see “Swivel Mount” on page 65
for additional details)
. When mounting the probe holder frame, ensure the
attachment knobs
(Fig. 254)
are at the front (non crawler side).
1
2
Fig. 254 - Position primary and secondary probe holders
Fig. 255 - Align swivel mount with scan surface