PAGE 115 of 177
CX0344 Rev 05.3
5.10.3.1 Mounting a Pivoting Probe Holder Frame
1.
Attach the wedges that are to be used with the probe holders
(see “Probe
Holder Setup” on page 69 for additional details).
2.
Affix the probe holders
(with attached wedges)
to the probe holder frame. Place
the secondary probe holders at the front of the
1
frame while placing the
primary probe holders at the rear of the
2
frame system
(Fig. 266).
TIP:
Phased array wedges are designed to be pulled along a scan surface.
3.
Mount the pivoting probe holder frame to the crawler
(see “Swivel Mount” on
page 65 for additional details).
WARNING!
FALLING OBJECT HAZARD.
The Pivoting Probe Holder Frame is to be mounted
only in the right hand swivel mount. Mounting it in both
the left hand and right hand swivel mounts may cause
the crawler to fall and
SEVERE INJURY
or
DEATH
could result.
2
1
Fig. 266 - Connect frame to right drive module