PAGE 116 of 177
5.10.3.2 Pivoting Probe Holder Frame Setup - Longitudinal Scanning
To prepare the pivoting probe holder frame for longitudinal scanning, follow
these steps:
NOTE:
The swivel mount must be in a horizontal position during
longitudinal scanning (see “Swivel Mount” on page 60).
1.
Loosen the pivot wing knobs at the centre of the frame system
(Fig. 267).
Lower the left side of the frame system to align with the tangent of the
scan surface. Tighten the pivot wing knobs.
WARNING!
FALLING OBJECT HAZARD.
When scanning in the longitudinal direction with the
Pivoting Probe Holder Frame, operation must be
limited to driving in the longitudinal direction only.
Only very slight corrective steering is permitted.
Excessive steering may cause the crawler to fall and
SEVERE INJURY
or
DEATH
could result.
Fig. 267 - Loosen pivot wing knobs