PAGE 55 of 177
CX0344 Rev 05.3
To configure the
NAVIC
system for scanning using dual drive modules with a
pivoting probe holder frame, follow these steps
(see “Pivoting Probe Holder
Frame” on page 114)
:
1.
Mount and connect the
6
umbilical to
4
right drive module of the
crawler.
2.
Attach a configured
3
pivoting probe holder frame
(see “Pivoting Probe
Holder Frame” on page 114)
to the crawler
(see “Swivel Mount” on page
60)
.
3.
Optional:
►
Attach the
1
tracker
(see “Tracker” user manual)
to the front of the
3
pivoting probe holder frame.
►
Connect the
2
auxiliary cable to the
6
umbilical’s 4-pin expansion
connector.
CAUTION!
DO NOT DISCONNECT
UNDER LOAD.
Shut off power before connection or
disconnecting. Permanent damage to electronics
could occur.