5.7.7.5 Probe Holder Left/Right Conversion
To reverse the probe holder, follow these steps:
NOTE:
To perform this operation the 1.5 mm hex wrench (Fig. 53) is required.
1.
Ensure the probe holder is in latched, upper position
(Fig. 148)
.
2.
Using the supplied 3 mm hex driver
(Fig. 50)
, unscrew the yoke pivot screw
and remove yoke
(Fig. 158)
.
3.
Loosen the probe holder arm adjustment knob and the arm clamp screw.
Slide the probe holder arms off the yoke
(Fig. 159)
.
4.
Flip the yoke 180° and reverse the probe holder arms
(Fig. 160)
.
5.
Place the pivot buttons on the inside of the probe holder arms
(Fig. 161)
using a
3/8 in wrench
(Fig. 51)
.
Fig. 158 - Unscrew yoke pivot screw
Fig. 159 - Remove probe holder arms
Fig. 160 - Flip yoke and reverse arms
Fig. 161 - Attach arms & move buttons