PAGE 59 of 140
DP0029 Rev 02.2
5.8.2.5 Probe Holder Left/Right Conversion
To reverse the probe holder, follow these steps:
NOTE:
To perform this operation the 1.5 mm hex wrench (Fig. 40) is required.
1.
Ensure the probe holder is in latched, upper position
(Fig. 120)
.
2.
Using the supplied 3 mm hex driver
(Fig. 37)
, unscrew the yoke pivot
screw and remove yoke
(Fig. 130)
.
3.
Loosen the probe holder arm adjustment knob and the arm clamp screw.
Slide the probe holder arms off the yoke
(Fig. 131)
.
4.
Flip the yoke 180° and reverse the probe holder arms
(Fig. 132)
.
5.
Place the pivot buttons on the inside of the probe holder arms
(Fig. 133)
using a 3/8 in wrench
(Fig. 38)
.
Fig. 130 - Unscrew yoke pivot screw
Fig. 131 - Remove probe holder arms
Fig. 132 - Flip yoke and reverse arms
Fig. 133 - Attach arms & move buttons