PAGE 76 of 140
5.9.3.1 Mounting a Pivoting Probe Holder Frame
1.
If attached, use the 3 mm hex
driver to remove the probe
holder pivot mount from the
pivoting probe holder frame.
2.
Attach the wedges that are to
be used with the probe
holders
(see Probe Holder
Setup on page 55 for
additional details).
1.
Affix the probe holders
(with attached wedges)
to the probe holder frame. Place
the secondary probe holders at the front of the frame
(Fig. 176-1)
while placing
the primary probe holders at the rear of the frame system
(Fig. 176-2).
TIP
: Phased array wedges are designed to be pulled along a scan surface.
2.
Mount the pivoting probe holder frame to the crawler
(see Swivel Mount on
page 44 for additional details).
Fig. 175 - Remove probe holder pivot mount
2
1
Fig. 176 - Connect frame to crawler’s swivel mount