PAGE 80 of 142
5.10.3.1 Mounting a Pivoting Probe Holder Frame
1.
If attached, use the 3 mm hex
driver to remove the probe
holder pivot mount from the
pivoting probe holder frame.
2.
Attach the wedges that are to
be used with the probe holders
(see Probe Holder Setup on page
59 for additional details).
3.
Affix the probe holders
(with attached wedges)
to the probe holder
frame. Place the secondary probe holders at the front of the frame
(Fig.
188-1)
while placing the primary probe holders at the rear of the frame
system
(Fig. 188-2).
TIP:
Phased array wedges are designed to be pulled along a scan surface.
4.
Mount the pivoting probe holder frame to the crawler
(see Crawler on
page 46 for additional details).
Fig. 187 - Remove probe holder pivot mount
2
1
Fig. 188 - Connect frame to crawler’s swivel mount