Important Precautions 

Warning: Risk of Serious Injury or Death. 

These 

products are intended for use with Joerns

®

 bed models 

B530, B330, FreeStyle

®

, Easy Care

®

 2000-2004, 

Easy Care 3, 5, 7, 9, Care 100 and EasyCare. Use of 

these products on any bed they were not designed for 

could result in an unproven or unsafe configuration, 

potentially resulting in serious injury or death. For 

questions regarding compatibility, contact your Joerns 

representative.

Warning: Risk of Serious Injury or Death.

 Use a 

properly sized mattress in order to minimize the gap 

between the side of the mattress and the Vari-Care 

assist device or side rail. This gap must be small 

enough to prevent a resident from getting his/her head 

or neck caught in this location. Make sure that raising 

or lowering the bed, or adjusting the sleep surface, 

does not create hazardous gaps. Failure to do so 

could result in serious injury or death.

Warning: Risk of Serious Injury or Death.

 Properly 

locate the mounting brackets. The gap between the 

head/foot panel and the Vari-Care assist device or side 

rail must be small enough to prevent a resident from 

getting his/her head or neck caught in this location 

(see the installation instructions for more information, 

if applicable). If multiple assist devices are needed, 

position them such that the gap between them is 

large enough that the trunk and hips can easily pass 

through. Make sure that raising or lowering the bed, or 

adjusting the sleep surface, does not create hazardous 

gaps. Vari-Care assist devices or side rails should not 

be used if ANY openings within the bed system allow 

a resident to get his/her head or neck lodged within 

these openings. Failure to do so could result in serious 

injury or death.

Warning: Risk of Serious Injury or Death.

 Do not 

use the Vari-Care assist device as an assist until 

you have verified that it is up AND locked. Injury to 

resident or caregiver may result if this procedure is not 

followed.

Warning: 

An optimal bed system assessment should 

be conducted for each resident by a qualified clinician 

or medical provider to ensure maximum safety of 

the resident. The assessment should be conducted 

within the context of, and in compliance with, the 

state and federal guidelines related to the use of 

restraints and bed system entrapment guidance, 

including the 

Clinical Guidance for the Assessment 

and Implementation of Side Rails

 published by the 

Hospital Bed Safety Workgroup of the U.S. Food 

and Drug Administration. Further information can be 

obtained at the following web address: http://www.fda.

gov/MedicalDevices/ProductsandMedicalProcedures/

GeneralHospitalDevicesandSupplies/HospitalBeds/

default.htm

Maintenance/Inspection Information:

 Visually 

inspect the assist handle and mounting bracket, 

and check for loose hardware on a monthly basis. 

Tighten loose hardware as stated in the installation 

instructions.

Warning: 

In the event worn or broken part(s) are 

identified, remove the device from service until repairs 

are made.

Note: 

The Vari-Care assist devices are intended for 

use as an aid in entering or exiting the bed sleep area, 

as well as a stable handhold during self positioning 

within the sleep area.

Note: 

The Vari-Care side rails are intended to provide 

a physical indication of the bed edge, as well as a 

stable handhold during self positioning within the sleep 

area.

Joerns assist devices or side rails are designed to be 

used with specific Joerns bed frames. Once mounted 

on the appropriate bed frame, assist devices or 

side rails become an integral part of the Joerns bed 

system. As such, these mounted assist devices or side 

rails can handle the same safe working load, within 

normal use conditions, of the appropriate bed to which 

they are mounted. 

When assessing the risk for entrapment, you need to 

consider your bed, mattress, head/foot panels, assist 

devices and other accessories, as a system. It is also 

extremely important to review the resident’s physical 

and mental condition and initiate an appropriate 

individual care plan to address entrapment risk. All 

Joerns bed systems are evaluated for compliance to 

the FDA/CDRH "Hospital Bed System Dimensional 

and Assessment Guidance to Reduce Entrapment."

Summary of Contents for Vari-Care F14SC

Page 1: ...User Service Manual Joerns Assist Device and Side Rails Vari Care Models To avoid injury read user s manual before using...

Page 2: ...lowed Warning An optimal bed system assessment should be conducted for each resident by a qualified clinician or medical provider to ensure maximum safety of the resident The assessment should be cond...

Page 3: ...lines recommend dimensional limits for critical gaps and spaces between bed system components Entrapment zones involve the relationship of components often directly assembled by the healthcare facilit...

Page 4: ...6110015 RevJ 14 2315 Matching Side Rail Options to your Beds F16C F168 F1576C for 76 beds 58 00 15 00 11 25 IN RAIL BODY 4 41 LARGEST GAP OF SLEEP DECK APPROX 4 30 TO TOP 67 00 11 25 IN RAIL BODY 4 4...

Page 5: ...Combination For older model beds contact product service Joerns Healthcare recommends the use of a support surface with minimum dimensions of 35 in width by 6 in depth Support surfaces used in conjunc...

Page 6: ...he bed F14SC For B530 B330 and EasyCare 2000 2004 Note Latch must face the center of the bed F1580C F1576C F16C for 80 EasyCare 2000 2004 Note Latch must face the foot end of the bed F1576C For 76 Eas...

Page 7: ...w 4 6 F168 Not included in N832 Full Rail Mount Kit Shown for reference only Suggested tools 2 wrench or socket Installation Instructions 1 Raise the back and foot sections of the bed as shown in Figu...

Page 8: ...ustrations When complete there should be an alignment label on each side of the frame aligned with the cross member located closest to the head panel 4 Figure 4 shows a general assembly instruction Be...

Page 9: ...tighten the brackets to the frame until the next step 6 Position the head end brackets over the location labels Ensure that the edge of the bracket is flush with the arrow and is covering the black st...

Page 10: ...ent s hand slipping on the assist device and result in injury or death Warning In the event that a ground wire is disconnected during installation RECONNECT IT NOW Maintenance Inspection Information V...

Page 11: ...from the bed to the side rail Figure 20 Figure 21 Figure 22 A B C D E F G H Figure 19 Maintenance Inspection Information Visually inspect the assist handle and mounting bracket and check for loose har...

Page 12: ...e or defects caused by misuse incorrect handling or installation by a non authorized installer Damage caused by use in unsuitable environmental conditions or failure to maintain the product in accorda...

Reviews: