MAINTENANCE
Cleaning the Compressor’s Outer Case

 WARNING: Electrical shock hazard. Do not remove 

the outer case of the device. All disassembly and 
maintenance of the device must be performed by a 

qualified service technician. Refer service to qualified 

service personnel.

 WARNING: This device does not require oil. Do not 

attempt to lubricate any internal parts.

 WARNING: Unplug device before cleaning. Do not 

submerge in water for cleaning.

Wipe the compressor's outer case with a clean, damp cloth 
weekly to keep it dust-free. Do not use any powdered type 
cleaners or soap.

Cleaning the Nebulizer Cup

All parts of the disposable nebulizer, except tubing, should 
be thoroughly cleaned after each use. If your physician 
specifies a specific cleaning procedure, follow that 
procedure.

 WARNING: To prevent possible risk of infection from 

contaminated medication, clean the nebulizer after each 
treatment.

1.  After each use, disassemble the nebulizer cup, T 

adapter and mouthpiece completely, wash all parts in 
warm soapy water, and rinse thoroughly. Allow to air 
dry. Store in a clean, dry location.

2.  Once daily, after washing as above, soak nebulizer 

components in a fresh solution consisting of one part 
warm water and one part white vinegar for 30 minutes 
followed by a warm water rinse. Allow to air dry. Store 
in a clean, dry location.

Cleaning the Drawing Board

Wipe the drawing board outside cover with a clean, damp 
cloth weekly. Do not use any liquid, cleansers or soap on 
any surface of the drawing board.

Summary of Contents for Neb-u-Tyke Neb-a-Doodle JB0112-070

Page 1: ...rence Federal Law USA restricts this device to sale by or on the order of a physician This device should not be used unless the operator has been instructed by a qualified healthcare professional JB0112 070 INS LAB RevB14 Neb u Tyke Nebulizer Compressor Item JB0112 070 User Manual ...

Page 2: ...5 STORAGE 15 TROUBLESHOOTING 16 SPECIFICATIONS 17 LIMITED WARRANTY 18 COMPRESOR DE NEBULIZADOR MANUAL USUARIO EN ESPAÑOL 21 This product has been tested and successfully approved with the following standards UL1431 EN 60601 1 GF Health Products Inc is not responsible for typographical errors All illustrations specifications packaging and warranties contained in this literature are based on the lat...

Page 3: ...e of compressed air for aerosol therapy It is used in conjunction with a jet pneumatic nebulizer to produce medicated aerosols for inhalation by pediatric and adult patients with respiratory symptoms Indications for therapy include asthma chronic bronchitis infection of the upper respiratory tract chronic obstructive pulmonary disease and other respiratory disorders in accordance with a physician ...

Page 4: ...ctric shock for type BF equipment not applicable for UL1431 Disposal of Electrical Electronic Equipment WEEE This product should not be treated as household waste but submitted to an appropriate collection point for the recycling of electrical and electronic equipment For more detailed information about the recycling of this product please contact your local city office household waste disposal se...

Page 5: ...ld result in minor or moderate personal injury s NOTICE Indicates a potential hazard or unsafe practice that if not avoided could result in product or property damage Info Provides application recommendations or other useful information to ensure that you get the most from your product READ ALL INSTRUCTIONS BEFORE USING DANGER To reduce the risk of electrocution 1 Always unplug this product immedi...

Page 6: ...te this product if a it has a damaged cord or plug b it is not working properly c it has been dropped or damaged or d it has been dropped into water Return the product to your Graham Field authorized distributor for examination and repair 6 Keep the cord away from heated surfaces 7 Never block the air openings of this product or place it on a soft surface such as a bed or couch where the air openi...

Page 7: ... nebulizer 15 This device is oil free Do not lubricate 16 Do not remove the jet nebulizer while the device is operating 17 Do not leave medication inside the jet nebulizer when it is not in use 18 Use the nebulizer intermittently the maximum operating period is 30 minutes The nebulizer must stand unused for 30 minutes between uses 19 The mouthpiece and masks are for use by one person only in order...

Page 8: ...in the illustration below reservoir tube ON OFF switch shapes stylus slide tab drawing board nebulizer housing push button air outlet connector for tubing filter cap adult mask nebulizer cup mouthpiece tubing T adapter pediatric mask Neb u Tyke Neb a Doodle Nebulizer Compressor Info Neb a Doodle replacement parts are available ...

Page 9: ...ure you can easily reach the controls when seated 3 Open the nebulizer medication cup Twist the cap counter clockwise until it stops then lift cap off 4 Pour the prescribed medication into the nebulizer medication cup Do not fill with more than 5 ml per treatment 5 Install cap on cup Turn clockwise to tighten 6 If using the mouthpiece for treatment proceed to step 7 If using a mask install the bot...

Page 10: ...of the T adapter 8 Install T adapter onto top of nebulizer medication cup cap 9 Connect the tubing to the adapter on the bottom of the nebulizer medication cup 10 Connect the tubing s other end to the compressor air outlet connector WARNING Before use check that all parts are securely connected Use of an improperly assembled nebulizer could diminish or prevent adequate delivery of medication and t...

Page 11: ... the ON OFF switch to the ON position to start the compressor 4 Place the nebulizer mouthpiece between teeth and close mouth or install the mask on face 5 As aerosol starts flowing inhale deeply and slowly through mouthpiece or mask WARNING Never use a nebulizer with a clogged nozzle If the nozzle is clogged the aerosol mist will be reduced altered or stopped which will diminish or eliminate the e...

Page 12: ...ensures the drawing board stays in place To prevent damage to the unit ensure you do not insert it from the other direction To draw on the drawing board move the included stylus pen pointed end down along the drawing board To use the included shapes place them on the drawing board and press lightly To erase the drawing board move the slide tab over the drawing board WARNING This product contains s...

Page 13: ...l automatically shut off the motor Should this occur move the ON OFF switch to the OFF position and allow the motor to cool down for approximately 30 minutes before turning it on again 2 If the overload protector shuts off the motor frequently you may have an unstable voltage situation 3 If the device shuts down and cannot restart it may need to be replaced Call your Graham Field authorized distri...

Page 14: ...lizer except tubing should be thoroughly cleaned after each use If your physician specifies a specific cleaning procedure follow that procedure WARNING To prevent possible risk of infection from contaminated medication clean the nebulizer after each treatment 1 After each use disassemble the nebulizer cup T adapter and mouthpiece completely wash all parts in warm soapy water and rinse thoroughly A...

Page 15: ...he ON OFF switch 2 Remove the dirty filter with a small pointed object Discard the filter 3 Replace with a clean filter 4 Replace the filter cap s NOTICE Do not use cotton or other materials as a filter s NOTICE Do not wash or clean the filter s NOTICE Do not reuse the filter Do not operate compressor without a filter Compressor damage can result WARNING The disposable nebulizer kit is for single ...

Page 16: ...hen device has been run continuously for more than 30 minutes just before using auto shut down may have been activated by the built in thermal protector ensure that device is OFF for at least 30 minutes before next usage Nebulization is weak Check for proper electrical voltage Check tubing for blockage or air leakage at connection to Neb u Tyke compressor or nebulizer cup replace as needed Check t...

Page 17: ...rons Capacity 5 ml Filter Sponge Dimensions LxWxH 10 6 5 4 2 255 mm x 166 mm x 108 mm Weight 4 lb 1860 g Electrical requirement 115VAC 60Hz Current 1 2A rate Fuse 2 5A 250V Power cord Black 7 5 ft 90 2 286 mm Plug 2 prong type Sound level 56 5 dBA Environment Temperature Operation 41 F to 104 F 5 C to 40 C Storage 40 F to 158 F 40 C to 70 C Humidity Operation 30 to 95 RH Storage 10 to 95 RH ...

Page 18: ...shall commence from date of shipment to the original customer unless there is an expiration date on the component in which case the warranty shall expire on the earlier of warranty period or the expiration date EXCLUSIONS The warranty does not cover and GF shall not be liable for the following 1 Defects damage or other conditions caused in whole or in part by misuse abuse negligence alteration acc...

Page 19: ... DOWNTIME COVER OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES PAYMENTS AND BENEFITS The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understandings with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any rep...

Page 20: ...Manufactured for GF Health Products Inc Made in China ...

Reviews: