18 DePuy Synthes Instructions for Use Air Pen Drive
Attachments
Burr Attachments
Burr Attachments
(05.001.045–05.001.050, 05.001.055, 05.001.063)
Gear ratio: 1:1
The system includes straight and angled Burr Attach-
ments in 3 lengths each (S, M, L). The related burrs are
also marked with S, M and L. Angled XL and XXL Burr
Attachments are available; for these attachments the L
burrs shall be used.
Changing burrs
1. Lock unit.
2. Turn the release sleeve for burrs
1
until it engages in
the UNLOCK position and remove the tool.
3. Insert the new tool as far as possible, turn it slightly
until it locks in place and then turn the release sleeve
for burrs into the LOCK position until it engages. The
burr is correctly clamped when the marking S, M or L
on the burr shank is no longer visible.
Information on handling burrs
•
Synthes recommends using a new sterile burr for each
operation. This prevents health risks to the patient.
•
Used burrs present the following risks:
•
Necrosis due to excess heat
•
Longer cutting time due to reduced performance
of the burr
Precautions:
•
Burrs must be cooled with irrigation liquid to
prevent heat necrosis.
•
The size of the burr attachment must correspond
to the size of the burr (e.g. attachment size S with
burr size S) or one burr size bigger (e.g. attachment
size S with burr size M).
•
User and OR personal must wear safety goggles
when working with burrs.
•
When the Burr Attachments are not attached to
the handpiece during tool change, use the Hand-
hold (05.001.074) for easier change of burrs.
1
Release sleeve for burrs