Air Pen Drive Instructions for Use DePuy Synthes 27
Important:
•
Follow the instructions from chapter ”Preparation
Prior To Cleaning” before starting the manual
cleaning.
•
Do not clean the foot switch following the Manual
Cleaning Instructions.
1.
Remove debris.
Rinse the device under running cold
tap water for a minimum of 2 minutes. Use a sponge,
soft lint-free cloth or soft-bristled brush to assist in re-
moving gross soil. For cannulations of attachments,
the cleaning brush (05.001.075) shown below, should
be used.
Note:
Do not use pointed objects for cleaning.
Brushes shall be inspected before daily use and
discarded if they have degraded to the point
where they may scratch instrument surfaces or be
ineffective due to worn or missing bristles.
2.
Manipulate moving parts.
Manipulate all moving
parts such as the triggers, sleeves and switches under
running tap water to loosen and remove gross debris.
3.
Spray and wipe.
Spray and wipe the device using a
neutral pH enzymatic solution for a minimum of
2 minutes. Follow the enzymatic detergent manu-
facturer’s directions for correct temperature, water
quality (i.e. pH, hardness) and concentration/
dilution.
4.
Rinse with tap water.
Rinse device with cold tap
water for a minimum of 2 minutes. Use a syringe or
pipette to flush lumens and channels.
5.
Clean with detergent.
Clean the device manually un-
der running warm water using an enzymatic cleaner
or detergent for a minimum of 5 minutes. Manipulate
all moving parts under running water. Use a soft-bris-
tled brush and/or soft lint-free cloth to remove all visi-
ble soil and debris.
Care and Maintenance
Cleaning and Disinfection
Manual Cleaning Instructions
Cleaning Brush (05.001.075)