Important:
•
Follow the instructions from chapter ”Preparation
Prior To Cleaning” before starting the automated
cleaning with manual pre-cleaning.
•
Manual pre-cleaning prior to automated cleaning/
disinfection is important to ensure that cannula-
tions and other difficult to access areas are clean.
•
Alternative cleaning/disinfection procedures other
than in the procedure described below (including
manual precleaning) have not been validated by
Synthes.
•
Do not clean the foot switch following the Auto-
mated Cleaning Instructions with Manual Pre-
cleaning.
1.
Remove debris.
Rinse the device under running cold
tap water for a minimum of 2 minutes. Use a
sponge, soft lint-free cloth or soft-bristled brush to
assist in removing gross soil. For cannulations of the
handpiece and attachments, the cleaning brush
(05.001.075 shown below) should be used.
Note:
Do not use pointed objects for cleaning.
Brushes shall be inspected before daily use and
discarded if they have degraded to the point
where they may scratch instrument surfaces or
be ineffective due to worn or missing bristles.
2.
Manipulate moving parts.
Manipulate all moving
parts such as the triggers, sleeves and switches un-
der running tap water to loosen and remove gross
debris.
3.
Spray and wipe.
Spray and wipe the device using a
neutral pH enzymatic solution for a minimum of
2 minutes. Follow the enzymatic detergent manu-
facturer’s directions for correct temperature, water
quality (i.e. pH, hardness) and concentration/
dilution.
Cleaning Brush (05.001.075)
Care and Maintenance
Cleaning and Disinfection
Automated Cleaning Instructions
with Manual Pre-cleaning