32 DePuy Synthes Instructions for Use Air Pen Drive
9.
Automated cleaning cycle parameters
Note:
The washer/disinfector should fulfill the
requirements as specified in ISO 15883.
Step
Duration
(minimum)
Cleaning instructions
Rinse
2 minutes
Cold tap water
Pre-wash
1 minute
Warm water (≥ 40 °C); use detergent
Cleaning
2 minutes
Warm water (≥ 45 °C); use detergent
Rinse
5 minutes
Rinse with de-ionized (DI) or purified water (PURW)
Thermal disinfection
5 minutes
Hot DI water, ≥ 90 °C
Dry
40 minutes
≥ 90 °C
10.
Inspect the device.
Remove all the devices from the
washing basket. Inspect the cannulations, coupling
sleeves, etc. for visible soil. If necessary, repeat the
manual pre-clean/automated cleaning cycle. Confirm
that all parts are completely dry. If smaller devices or
cannulations contain residual water, blow dry with
medical grade compressed air.
Automated cleaning/disinfection is an additional
stress for power equipment, especially for seals and
bearings. Therefore, systems must be properly lubri-
cated and regularly sent to be serviced (at least once
per year).
Care and Maintenance
Cleaning and Disinfection
Automated Cleaning Instructions with Manual Pre-cleaning