1
3
2
Air Pen Drive Instructions for Use DePuy Synthes 35
Maintenance of movable parts on handpiece
and Foot Switch – with Synthes Maintenance Oil
(05.001.095)
Apply Synthes Maintenance Oil 05.001.095 for Electric
and Air Pen Drive after each use or when needed, on
movable parts on the handpiece and the Foot Switch.
Oiling movable parts of the handpiece
Apply one drop of oil in the slots between the adjust-
ment sleeve
1
and the basic body, one drop of oil in the
slots behind the release sleeve
2
and move the sleeves.
To oil the valve
3
of the handpiece first set the hand-
piece on Hand Switch position. Then drop one drop of
oil on the valve and connect the Hand Switch to the
handpiece. Operate the Hand Switch by moving it up
and down. Make sure that the valve opens and closes
and that oil get into the valve. If necessary, repeat the
oiling procedure. Finish by wiping off excess oil from the
handpiece.
Oiling movable parts of the Foot Switch
If the connections for air hoses and irrigation do not
move smoothly please drop one drop of Synthes Mainte-
nance Oil 05.001.095 on each connection and move
them to spread the oil.
Precaution:
Only use the Synthes Maintenance
Spray (05.001.098) or/and Synthes Maintenance Oil
for Electric and Air Pen Drive (05.001.095). Their
biocompatible composition matches the require-
ments for power tools in the operating room. Lubri-
cants with other compositions may lead to sticking
and could have a toxic effect.