6 DePuy Synthes Instructions for Use Air Pen Drive
Air Pen Drive System
Air Pen Drive
Mounting the air hose on the handpiece
The Air Hose (05.001.083 or 05.001.084) is coupled by
fitting the pins on the hose coupling into the grooves on
the coupling for hoses on the handpiece and turning the
hose coupling clockwise. Attach the other end of the air
hose to the compressed air or nitrogen source or to the
Foot Switch (05.001.081, see page 10). Ensure that the
coupling geometry of the air hose is compatible with the
geometry of the wall coupling. If the operating room
does not have an exhaust air system then use the Air
Diffusor (519.950) to diffuse the air. The Air Diffusor is
connected between the source and the air hose. To dis-
connect the air hose simply turn the hose coupling piece
counterclockwise and pull it off the handpiece. Remove
the air hose from the compressed air or nitrogen source
or from the Foot Switch.
Adjusting the handpiece
1 Adjustment sleeve
2 Release sleeve for attachment
3 Coupling for hoses
4 LOCK position
5 Hand Switch position
6 Foot Switch position
7 Lock slide for adjustment sleeve