DINAMAP*
Compact
Vital Signs Monitor
33
Selection of the
Remote Operation
check box toggles remote operation on and off. When on, a tick
appears next to the box, denoting that remote operation (i.e. from the central station) is possible.
Selection of the
Address
box allows the monitor’s address to be set within the range 1 to 255.
Selection of
OK
returns to the Service Menu.
System
Selection of this button causes the hardware System menu to appear.
Selection of any of the check boxes causes that monitor feature to be toggled on or off. When on, a
tick appears next to the box, denoting that the option is available. Monitors without certain features are
supplied without the necessary hardware for that feature and thus will have that option unselected.
Selection of
OK
saves changes and returns to the Service menu.
OK
Selection of this button returns the display to the Service Menu.
3.2.2.5 Silence
Selection of this button will cause
all
patient
alarms to be muted until either the monitor is powered off
and on again, or the Alarm Silence button is pressed.
A confirmation menu will appear on the display. Selection of either
Yes
or
No
will exit the menu.
If silence is confirmed, the monitor’s Alarm Silence button illuminates.
Sp0
2
√
X
Temp:
Printer
√
OK
√
X
√
X
√
√
√
No
Yes
Confirm Silence