DINAMAP*
Compact
Vital Signs Monitor
35
Bulb Pump
(or Syringe)
Airtight container (200-500ml recommended)
(Not necessary with syringe)
Manometer
DINAMAP* COMPACT
Calibration check or re-calibration
3.3.2 NIBP Calibration
The following section describes how to perform NIBP re-calibration on a DINAMAP*
Compact
Vital
Signs Monitor. Refer also to the previous section for details on checking of the DINAMAP*
Compact
calibration.
NIBP calibration is possible without disassembling the DINAMAP*
Compact
.
I. Set up the DINAMAP*
Compact
and calibration equipment as shown on page 35.
II. Switch on the DINAMAP*
Compact
and allow 10 seconds to settle.
III. Verify that the calibration pressure measurement system reads 0 mmHg (0 mBar)
IV. If the DINAMAP*
Compact
displays the language choice menu, choose “English”, then
“OK”. Reboot the DINAMAP*
Compact
by turning monitor off, then back on.
V. Select the “Mode” button from the main menu. The Mode menu will be displayed
VI. Select the “Service” button from the mode menu
VII. Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
VIII. Select “NIBP” from the service menu. The NIBP Service menu will be displayed.