DINAMAP*
Compact
Vital Signs Monitor
46
Disconnect pnuematic hose from pump assembly to Main Board secondary pressure transducer on
revision 06 units. (Disconnect pressure sensor cable on earlier revisions)
The boards can now be slid together up and out of the pcb guides built into the metalwork of the
chassis.
To separate the boards, use a fine nosed pliers to pinch the tip of the nylon p.c.b. spacer while gently
pulling the boards apart. Note that this will result in the loss of the user settings.
To avoid damage to the p.c.b. interconnection, ensure that the boards are not tilted or twisted relative
to each other as they are separated.