DINAMAP*
Compact
Vital Signs Monitor
49
3.4.7 Reassembling the Monitor
Re-assembly is a direct reversal of disassembly, however additional care must be taken to ensure that
all cables and hoses are refitted correctly. The main points to note when reassembling are:
•
Pay particular attention to the wire dressing and ensure that all ferrite components are refitted
correctly.
•
When reassembling the chassis, remember to reconnect the valve control from the pump to
the power supply board.
•
When refitting the front panel, ensure that the copper earthing fingers of the front panel
engage correctly inside the rear enclosure.
•
Ensure that the hoses to the front panel are not trapped or kinked while refitting the panel.
•
Keep your work area clean. The plastic display panel cover may become scratched if the
front panel is pressed against a hard object such as a screw.
•
Be sure to re-calibrate and test the monitor as per section 3.3 and 4 before releasing it
for clinical use.