DINAMAP*
Compact
Vital Signs Monitor
51
4.2 Test Procedure
4.2.1 Power Up Tests
4.2.1.1 LED Display
Press and release the main power on switch.
Check that all LED segments are illuminated momentarily.
Press and release the main power on switch.
Ensure unit shuts down.
4.2.1.2 LCD Display
Press and release ROTOR push button.
Confirm that all the Pixels of the LCD display are lit momentarily and that a series of three long beeps
followed by three short beeps can be heard.
Confirm that the display indicates the UUTs software revision momentarily.
If the instrument is Uncalibrated the LCD will display an NIBP calibration request prompt.
Press and release main power switch.
If the UUT requires calibration refer to section 3.3 of the Service Manual.
4.2.2 External DC Supply and Battery Indication Operation
4.2.2.1 UUT Supply Indicator Test
Switch on UUT wait for UUT to initialise and confirm that the External DC LED is lit and the battery LED
is extinguished.
Remove Battery and confirm that External DC LED on UUT Flashes.
Fit Battery and switch off External DC supply.
Confirm the External DC LED has extinguished and the Battery LED is lit.
Switch on External DC supply.
4.2.3 Temperature Requirements
Procede to the next section if temperature circuitry has already been calibrated and checked according
to section 3.3.5 and 3.3.6
If TEMP option is unavailable on main operational menu of the UUT LCD.
Enter Service Mode as detailed in section 4.1.3.1
Using rotor switch select SYSTEM from sub menu
Select SYSTEM again from next menu options
Select TEMP option box on UUT LCD, press rotor push button and confirm it has a tick against it.
Switch off UUT.