Quick Start Up Guide
RE-8500
6
Jovy Systems® Limited
4-
Standard profile operating method
The standard profile mode behavior is controlled; according to the IPC
J-STD-20 recommendations. Heaters power level controlled as well and
operating sequences between the upper heater, middle heater and side
heaters.
Profile is only running when two thermocouples are connected,
if one thermocouple connected process will not start.
Thermocouple (TC1) should place near the targeted component
and thermocouple (TC2) placed underside the PCB below the
component.
The table below shows the temperature ranges for each
profile stage in both Leaded and Lead Free profiles.
(Table -1)
Leaded Profile Lead Free Profile
Pre-heat Temperature
Range
90◦C to 135◦C
120◦C to185◦C
Soaking Temperature
Range
140◦C to 180◦C
190◦C to 215◦C
Peak Temperature Range
195◦C to 245oC
220◦C to 270◦C
Maximum Dwell time
250 seconds
250 seconds
Summary of Contents for RE-8500
Page 1: ...Quick Start Up Guide RE 8500 Jovy Systems Limited...
Page 64: ...Quick Start Up Guide RE 8500 64 Jovy Systems Limited 3 4 a PID UH MH SH Change RUN TAL...
Page 73: ...Quick Start Up Guide RE 8500 73 Jovy Systems Limited 1 Ok 2 Ok 3 Ok 4 5 6 7 8 9...
Page 77: ...Quick Start Up Guide RE 8500 77 Jovy Systems Limited 3 C Process Data File UH MH SH TAL...
Page 80: ...Quick Start Up Guide RE 8500 80 Jovy Systems Limited Leaded Lead free PID Profile Editor Tab...
Page 81: ...Quick Start Up Guide RE 8500 81 Jovy Systems Limited 1 230V 16A 3 6kw...
Page 85: ...Quick Start Up Guide RE 8500 85 Jovy Systems Limited 1 1 2 1 3 1 4 1 5 6 7 8 1 55 9 1 9 Z A...
Page 87: ...Quick Start Up Guide RE 8500 87 Jovy Systems Limited 2 1 2 1 2...
Page 88: ...Quick Start Up Guide RE 8500 88 Jovy Systems Limited 1 55 1 9 Z A...