The status symbol will change to a
–
when the error is not active, however, the
N
indicates that the alarm condition was not acknowledged. This allows you to
determine the error and alarm even if the error is no longer active.
5.6
Replacement of the Electrode
When the red (LED) "Alarm" is illuminated and the alarm message number 7 is
shown ‘Ag depleted” the silver electrode must be replaced. The electrode will give
you warnings at 80%, 85%, 90% and 95% depleted.
These errors can be
acknowledged and the system will continue to operate. This gives you sufficient time
to order a replacement electrode. Once the “Ag depleted” error message is shown
the unit will stop operating.
1. Replace the electrode by removing the black plug by pulling up. Care must
be taken to pull on the body of the plug and not the wire.
2. Shut off the ball valves before
and after the chamber. (As
shown)
3. Remove the electrode by
unthreading it from the
chamber.
4. Check that the chamber is
clean and free from foreign
objects and if necessary clean
the chamber.
5. Place the new electrode inside
the chamber making sure that
the provided O-ring is properly
seated.
6. Thread the new electrode into
the chamber and reconnect the
plug.
7. The unit must be told that the
electrode was replaced so that
Error Screen with an error no longer
active that has not been acknowledged.