9
Error messages
Error 1
Pressure sensor failure or overpressure.
Error 2
PT100 probe failure or overtemp.
Error 3
External alarm. Pressure switch or safety thermostat trigered.
To reset the safety thermostat press its button number
23
The thermostat is adjusted in the factory it must not be ope-
rated except to reset it.
Error 4
Manual valve (16) opened during a cicle.
Error 5
Opened door during a cycle.
Error 6
Pressure over +0.05bar with the autoclave stopped.
Error 7
E2PROM failure.
Error 8
Out of limits temperature during the sterilization.
Error 9
Mains supply failure during cycle.
10
Spare parts
Presoclave II 80 litres. Part No. 4002136
07292 RS-232 communication wire
07315 26 pins communication wire
13006 Contactor
15411 Fuse 16 A
16055 Solenoid electromagnet
16057 Electrovalve NA 3 mm 24 V
16212 Pressure switch
20002 Lid microrruptor
20067 Valve microrruptor
20096 Bipolar switch
29358 Frontal circuit
29362 Power circuit
36027 Solid relay 25A
39001 Heating element 3680W/230V
43031 Thermostat
43071 Pt100 probe