8
KEEP THE BATERY FAR FROM THE FIRE ; DO NOT TRY TO RELOAD OR TO SHORT CIRCUIT IT. USE ONLY BATTERIES
SUPPLIED BY JRI (REF 10102)
1
3
4
2
5
V.
FEATURES
FEATURES
Mini SPYRF GREEN
Measurement range
-40 +85°C
Number of channels
1 internal or external probe
Remote sensor
Ø4 x50mm / 0,5m Cable
Type of input
PTC
Accuracy
±0,4°C from -20 à +30°C
±0,5°C from -40 to -20°C & +30 to +80°C
±0.7°C above +80°C
Recording interval
1sec to 90 min
Memory size
10 000 measur1ko
Operating conditions
-30 +70°C
Temperature for storage
-40 + 85°C
Radio range (in free field)
Max 58 m adjustable
Radio band
868MHz or 902MHz
Battery lifetime
2 years
Dimensions
85x55x18mm
Protection level
IP67
EN 12 830 compliance
Yes : this device must be regularly checked according to
EN 13 486 (1 time per year recommended)
CE ERM compliance
EN 301 489 / EN 61000 / EN 61010
EN 55022 / EN 300 220
VI.
REPLACING BATTERY
When the Mini SPY RF Green battery has to be replaced, a technical alarm is processed by the software Sirius
To replace the battery, follow the instructions below :
Restart the device by pressing 5’’ on the pushbutton after changing the battery.
DOWNLOAD THE SPY RF MEMORY BEFORE CHANGING THE BATTERY. DO NOT WAIT
UNTIL THE BATTERY IS EMPTY OTHERWISE THE MEMORY WILL BE DELETED
Open the MiniSpy
RF Green
Replace the battery
Short circuit the two pins
with a paper clip to reset
the device
Replace the seal and
close the device