background image

©JRI 

 

14 

 

 

b)

 

Programmed Start Delay 

The  SPY  IP  displays  the  date/time  at  which  the  relevant  measuring  channel  will  start 
recording. 

Caution

: This mode is only valid once. If the recording is stopped the SPY IP will indicate 

that it is awaiting configuration. 

 
 
 
 
 
 

c)

 

Push-button 

The  temperature  display  screen  will  show  only  a  push-button  for  channels  being 
configured in this mode. 
 
 
 
 
 
When you press the push-button an identification window will become available. 
Enter  the  identification  code.  Once  approved

 

by  Sirius  the  relevant  measuring  channel 

will start recording. If the operator does not have the rights, the action will be rejected 
and  an  error  message  is  displayed  for  a  few  seconds  before  the  push-button  appears 
again. 

 

d)

 

Target value reached: 

This type of start mode can only be configured channel by channel. The channel will start 
recording  when  it  has  reached  its  target  value  taking  into  account  the  direction  of  the 
variation  (upward  or  downward  movement).

 

If  a  channel  has  no  target  value,  it  is 

considered as being configured in « immediate » start mode. 

 

 

ou 

 
 

 

e)

 

Dry contact (change of state) 

The  recording  of  one  channel  can  be  regulated  by  the  change  of  state  of  a  contact 
connected  to  one  of  the  other  channels.  The  screen  of  the  channel  waiting  to  start 
recording will look like one of the following figures: 

These  figures  show  that  channel  1  will  start 
recording  when  channel  2  changes  state 
(rising or falling edge) 
Note:  when  channel  2  returns  to  its  original 
state, channel 1 will stop recording. 

 

2)

 

Off Modes 

All  « off  modes »  are  compatible  with  all  «  start  modes  »except  for  « immediate  »  start 
mode because as soon as the stop has been requested the SPY IP will start recording again. 
The programmed « off mode » will appear in the bottom right-hand corner (zone 5) 
 
       or         Target value                   Programmed stop delay          Push-button 
        Memory Full 
 

Current 
instantaneous value 

Target value 

Summary of Contents for SPY IP 4

Page 1: ...USER MANUAL SPY IP 10486F...

Page 2: ...Start modes 10 c Zone 4 Alarm Thresholds 10 d Zone 5 Off Modes 10 e Main Menu 10 4 BACKLIGHT COLOURS 10 5 SCREEN NAVIGATION 11 a Main menu 11 b Menu Configuration 11 c Device Status Menu 12 d Alarm M...

Page 3: ...omplies with current regulations CE MARKING this device complies with EU legislation for electric security inflammability disturbing radiation emission and immunity to surrounding electric disturbance...

Page 4: ...guration information from the DHCP server If the UDP broadcast is not allowed you must carry out a manual installation on Sirius Storage using the IP address provided by the DHCP server and the serial...

Page 5: ...twork UDP packets are never transmitted and you must carry out a manual installation If the manual installation fails make sure that the internal serial number matches the serial number printed on the...

Page 6: ...3 Declare in Sirius measures RH on channels 2 and or 4 c Identification of the connectors d Wiring Rear view of the connectors Probe 1 to 4 Ethernet Network RJ45 Modular connector Male plug end on the...

Page 7: ...JRI recommends using the cable with the reference number 00949 for connecting equipment which is not provided with a compatible socket f Battery backup Minimum battery backup time is 6 hours with a 3...

Page 8: ...e alone Humidity Measurement range resolution accuracy probe alone 40 C to 85 C 0 01 C 0 2 C from 20 C to 50 C 0 4 C outside 0 100 0 01 2 from 20 to 80 3 5 outside Current input 0 20 ou 4 20 mA measur...

Page 9: ...avoid damaging the touchscreen do not use sharp instruments and do not press hard 2 Touchscreen Description Channel ID Status of the device Measurement and start modes Channel number and alarm thresh...

Page 10: ...resholds High Pre Alarm High Alarm Low Pre Alarm Low Alarm d Zone 5 Off Modes Push Button Programmed stop delay Target value upward movement Target value downward movement e Main Menu Back to previous...

Page 11: ...r a pencil tip See Principle above Press the screen to access the main menu Back to measurement display a Main menu Select the desired zone by pressing the relevant icon b Menu Configuration The SPY I...

Page 12: ...dgment from the touchscreen Spy IP The action will be authorized and stored by Sirius The password change will just be made from Sirius Default IP Spy has 1 Profile Administrator mdp 1111 V HOW THE SP...

Page 13: ...mmed via Sirius are Immediate Pushbutton Programmed start delay Target value Change of state of one of the other channels a Immediate default mode in Sirius The channels set with this start mode will...

Page 14: ...art recording when it has reached its target value taking into account the direction of the variation upward or downward movement If a channel has no target value it is considered as being configured...

Page 15: ...t got the rights the request will be rejected and the channel will continue recording d Target value This off mode can only be used and configured channel by channel The channel will stop recording w...

Page 16: ...re alarm then the backlight colour will be the same for all the channels according to the most important even In this case it will be RED The threshold alarms can be either instantaneous or continuous...

Page 17: ...hout prior notice IX MAINTENANCE CONTRACT With our maintenance contract you will benefit for a minimal period of 2 years from the following advantages material verification once or twice a year warran...

Reviews: