MAINTENANCE
24
Warning
: Power should be disconnected before beginning
preventive maintenance on the concentrator.
DO NOT service or maintain while patient in use.
The concentrator need no extra approved maintenance as it pressure
and oxygen purity self-check of the unit, so it but the device shall be
maintained once a year. In location with much dust the maintenance can
be performed if necessary. The minimum maintenance is the insurance of
oxygen concentrator.
5.1 Cleaning the Cabinet
Clean the cabinet at least once a month.
Turn off the power switch and unplug the concentrator’s AC connector from
the power outlet.
Only the outside of the concentrator is to be cleaned. Use a soft dry
cloth, a damp sponge or wipes with alcohol based solution. Do not
use acetone, solvents or any other inflammable products. Do not spill
liquid inside the cabinet.
5.2 Cleaning or Replacing the Filter (3 Types)
Clean and replace the filters as often as specified in the following
paragraphs in order to protect the compressor and extend the
concentrator’s life.
DO NOT operate the concentrator without the filters
installed , or filters are wet. These actions could permanently damage the
concentrator.