31
7 Startup
7.5
Setting the zero point (offset) (Off.P)
7.5.1
Automatic offset setting
This setting applies the current measured value as the new zero point.
NOTE!
Automatic offset setting is only possible for devices with relative pressure measuring range!
1. Unlock the device.
chapter 7.2 "Unlocking the device (enter the code).", Page 27
2. "Turn" until "Off.P" appears in the bottom line.
3. "Press" twice in quick succession.
This applies the current measured value as the zero point.
7.5.2
Edited offset setting
This setting lets you increase or reduce the measured pressure by a configurable value.
1. Unlock the device.
chapter 7.2 "Unlocking the device (enter the code).", Page 27
2. "Turn" until "Off.P" appears in the bottom line.
3. "Press"
NOTE!
"-" means: the offset is negative - the measured pressure is reduced.
The value is entered "digit-by-digit".
Summary of Contents for DELOS SI
Page 2: ...Further information and downloads qr 405052 en jumo info ...
Page 57: ......