10 Overcoming errors and malfunctions
10
Overcoming errors and malfunctions
Display
Possible cause
Measure
•
Measuring range overflow or un-
derflow
•
Probe break
Configuring another measuring
range, see chapter 7.4 "Select the
unit of measured (Uni.P)", Page 28
Device error:
•
1 = Internal
communication error
•
2 = Error, analog output
•
3 = Short circuit,
switching output 1
•
4 = Short circuit
switching output 2
•
5 = VCC 8 V outside the
working range
•
6 to 8 = Internal
communication error
•
9 = Invalid configuration
1, 6, 7, 8: Return device to Service;
see reverse of operating manual for
the address.
2: Check ambient temperature.
Check output for cable break. Output
load too high (for current output) or
too low (for voltage output).
3, 4: Check the corresponding switch-
ing output.
5: Check voltage supply.
9: Check the configuration.
Display overrun
Upper display: „ - - - -“
Lower display: Parameter name
Value is less than -9999 or greater
than +9999.
Check the corresponding switching
output.
Check configuration.