8 T
e
chnical
data
39
8 Technical
data
Input for thermocouple
1. These ranges refer to the ambiente temperature of 20°C
Input for standard signals
2.
0—1V, R
E
>10M
Ω
for Type 702040/41 with 2 relays
3.
0.2—1V, R
E
>10M
Ω
for Type 702040/41 with 2 relays
Designation
Range
1
Fe-Con
L
Fe-Con
J
EN 60584
Cu-Con
U
Cu-Con
T
EN 60584
NiCr-Ni
K
EN 60584
NiCrSi-NiSi
N
EN 60584
Pt10Rh-Pt
S
EN 60584
Pt13Rh-Pt
R
EN 60584
Pt30Rh-Pt6Rh B
EN 60584
-200 to + 900°C
-200 to +1200°C
-200 to + 600°C
-200 to + 400°C
-200 to +1372°C
-200 to +1300°C
0 — 1768°C
0 — 1768°C
0 — 1820°C
Measurement accuracy:
≤
0.4% / 100ppm/°C
Cold junction:
Pt100 internal
Designation
Range
Voltage
0 — 10V, R
E
> 100k
Ω
2
2 — 10V, R
E
> 100k
Ω
3
R
E
- input resistance
Current 4 — 20mA, voltage drop
≤
3V
0 — 20mA, voltage drop
≤
3V
Measurement accuracy:
≤
0.1% / 100ppm/°C
Input for resistance thermometer
Outputs
Relay:
n.o.(make) contact; 3A at 250V AC resistive load;
150,000 operations at rated load
Logic 0/5V:
Current limiting: 20mA; R
load
≥
250
Ω
Logic 0/12V:
Current limiting: 20mA; R
load
≥
600
Ω
Supply
110 — 240V AC -15/+10% 48 — 63Hz, or
20 — 53V AC/DC 48 — 63Hz, or
10 — 18V DC (Connect to SELV or PELV)
Designation
Range
Pt100 EN 60751
-200 to +850°C
Pt1000 EN 60751
-200 to +850°C
KTY11-6
-50 to +150°C
Measurement accuracy:
Pt100/1000:
≤
0.1% / 50ppm/°C
KTY11-6:
≤
1.0% / 50ppm/°C
Sensor lead
resistance: 20
Ω
max. per lead
Meas. current:
250µA
b70.2040en.bk Seite 39 Mittwoch, 19. November 2008 10:15 10