3 Operation and visualization
42
Instrument info
- This menu also includes the version number (e.g. 133.01.01), i.e. the
version number of the instrument software. It is important because some
functions are only available from a certain version number onwards.
- If the “Digital-I/O“ is available (Yes), then the recorder also contains an
RS232/RS485 interface.
Error
The following errors are possible:
Error
Description
none
Instrument o.k.
Data lost
A discharge of the battery/storage capacitor
occured during the last lengthy supply
interruption.
The clock was set to 01.01.97 00:00:00.
Remedy:
Reset the time (
v
Chapter 4.2.2 “Configuration -
Battery empty
This message appears on instruments with lithium
battery when the time was reset after a data loss.
Please return the instrument to the supplier for a
change of battery.
H
Data can be lost after separating the instrument from the supply
voltage after more than 10 years on instruments with a lithium bat-
tery, and after more than 2 weeks (15 — 25°C ambient tempera-
ture) on instruments with a storage capacitor.
Summary of Contents for LOGOSCREEN 500
Page 1: ...Paperless recorder B 95 5015 Operating Instructions 04 02 00378469...
Page 2: ......
Page 6: ...Contents...
Page 48: ...3 Operation and visualization 46...
Page 66: ...4 Configuration parameters 64...
Page 76: ...5 Setup program 74...
Page 87: ...85 10 T V Report on Data Manipulation Security...
Page 88: ...10 T V Report on Data Manipulation Security 86...
Page 89: ...87 10 T V Report on Data Manipulation Security...
Page 90: ...10 T V Report on Data Manipulation Security 88...
Page 97: ......
Page 98: ......
Page 99: ......