●    The device may not work for all patients. If you are unable to achieve 
stable readings, discontinue use. 

●    The update period of data is less than 5 seconds, which is changeable 
according to different individual pulse rate. 

●    The instrument does not have low-voltage alarm function, it only shows 
the low-voltage, please change the battery when the battery energy is used 
out. 

●    Batteries must be removed if the device is going to be stored for more 
than once month, or else batteries may leak. 

1.4 Indication for use 

The Fingertip Pulse Oximeter is a non-invasive device intended for the 
spot-check of oxygen saturation of arterial hemoglobin (SpO

2

) and the pulse 

rate of adult and pediatric patients in home and hospital environments 
(including clinical use in internist/surgery, anesthesia, intensive care etc.). 
This device is not intended for continuous monitoring. 

  2 Overview 

 
The pulse oxygen saturation is the percentage of HbO

2

 in the total Hb in the 

blood, so-called the O

2

 concentration in the blood. It is an important 

bio-parameter for the respiration. For the purpose of measuring the SpO

2

 

more easily and accurately, our company developed the Pulse Oximeter. At 
the same time, the device can measure the pulse rate simultaneously. 
The Pulse Oximeter features in small volume, low power consumption, 
convenient operation and being portable. It is only necessary for patient to 
put one of his fingers into a fingertip photoelectric sensor for diagnosis, and a 
display screen will directly show measured valued of Hemoglobin Saturation. 

2.1 Classification: 

Class II a (MDD 93/42/EEC IX Rule 10) 
Class II (U.S.FDA) 

2.2 Feature 

■    Operation of the product is simple and convenient. 

■    The product is small in volume, light in weight (total weight is about 50g 
including batteries) and convenient in carrying. 

■    Power consumption of the product is low . 

■    The product will automatically be powered off when no signal is in the 
product within 16 seconds. 

2.3 Major Applications and Scope of Application 

The Pulse Oximeter can be used to measure human Hemoglobin Saturation 
and pulse rate through finger, and indicate the pulse intensity by the 
bar-display. The product is suitable for use in family, hospital (Ordinary 
sickroom), Oxygen Bar, social medical organizations and also the measure of 
saturation oxygen and pulse rate. 
 

The product is not suitable for use in continuous supervision for patients. 

2.4 Environment Requirements 

Operation Temperature: 5

-40

 

Storage Temperature: -10

-50

 

Ambient Humidity: 15%-80% RH, no condensation in operation 
                                  10%-93% RH, no condensation in storage 
Atmospheric Pressure: 70 kPa to 106 kPa, in operation 

                                       

  50kPa

106 kPa, in storage 

  3 Principle and Caution 

 

3.1 Principle of Measurement 

Principle of the Oximeter is as follow: An experience formula of data process 
is established taking use of Lambert Beer Law according to Spectrum 
Absorption Characteristics of Reductive Hemoglobin (Hb) and 
Oxyhemoglobin (HbO

2

) in glow & near-infrared zones. Operation principle of 

the instrument is : Photoelectric Oxyhemoglobin inspection Technology is 
adopted in accordance with Capacity Pulse Scanning & Recording 
Technology, so that two beams of different wavelength of lights can be 
focused onto human fingertip through perspective clamp finger-type sensor. 
Then measured signal can be obtained by a photosensitive element, 
information acquired through which will be show on screen through treatment 
in electronic circuits and microprocessor. 

Summary of Contents for JPD-500E

Page 1: ...lified service engineers ONLY Users are not permitted to maintain it by themselves The oximeter cannot be used together with devices not specified in User Manual Only the accessory that appointed or recommendatory by manufacture can be used with this device This product is calibrated before leaving factory 1 2 Warning Explosive hazard DO NOT use the oximeter in environment with inflammable gas suc...

Page 2: ...eature Operation of the product is simple and convenient The product is small in volume light in weight total weight is about 50g including batteries and convenient in carrying Power consumption of the product is low The product will automatically be powered off when no signal is in the product within 16 seconds 2 3 Major Applications and Scope of Application The Pulse Oximeter can be used to meas...

Page 3: ...fluorescent lamp dual ruby light infrared heater direct sunlight and etc 4 Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy 4 Accessories 1 One hanging rope 2 Storage bag 2 Two AAA batteries 3 One user s manual 5 Installation 5 1 View of the Front panel Figure 2 Front View 5 2 Battery Step 1 Refer to Figure 3 and insert the two AAA size batteries...

Page 4: ... air or clean it by dry clean fabric Using the medical alcohol to disinfect the product after use prevent from cross infection for next time use Please take out the batteries if the oximeter is not use for a long time Warning High pressure sterilization cannot be used on the device Warning Do not immerse the device in liquid Warning It is recommended that the device should be kept in a dry environ...

Page 5: ...ormation Display Mode Display Format LED display The Pulse Oxygen Saturation SpO2 Digital Pulse Rate PR Digital Pulse Intensity bar graph Digital bar graph display SpO2 Parameter Specification Measuring range 35 100 the resolution is 1 Accuracy 70 100 2 Below 70 unspecified Pulse Parameter Specification Measuring range 25bpm 250bpm the resolution is 1 bpm Accuracy 2bpm Pulse Intensity Range Contin...

Page 6: ...dical com Postal Code 518103 DE Überblick Die Sauerstoffsä ttigung ist der Prozentsatz von Oxyhä moglobin HbO2 der mit Sauerstoff gegen alle kombinierbaren Hä moglobinen Hb kombiniert wird Es ist ein wichtiger physiologischer Parameter der in die Atmung und den Kreislauf einbezogen ist Mit diesem Instrument kann die Pulsfrequenz über die Arterie gemessen werden Produktmerkmale Leicht tragbar und e...

Page 7: ... von der Bildschirmanzeige 5 Drücken Sie im Boot Strap Zustand die Taste und das Gerä t wird zurückgesetzt Fingernä gel und Lumineszenzrohr sollten auf derselben Seite liegen Fehlersymptome Ursachen und Lösungen Symptom Möglice Ursache Solution Das Oximeter zeigt keine Daten zur Sauerstoffsä ttigung und Pulsfrequen z an 1 Der Finger befindet sich nicht in der richtigen Position 2 Der Patient konze...

Page 8: ...it Poids léger utilisation portable et opération simple Écran LED lisible Indication de basse tension de la batterie L appareil s éteindra automatiquement 16 secondes plus tard si aucun signal n est détecté L appareil est équipé de deux piles sèches 1 5 V AAA qui peuvent lui fournir de l énergie pour fonctionner plus de 24 heures vue du panneau avant 1 affichage basse tension 2 fréquence du pouls ...

Page 9: ...tacter le centre de service local L écran s éteint soudainement 1 L appareil s éteint automatiquement en l absence de signal dans les 16 secondes 2 Les batteries sont presque épuisées 1 Normal 2 Changez les piles IT Panoramica La saturazione di ossigeno è la percentuale di ossiemoglobina HbO2 che è combinata con l ossigeno contro tutta l emoglobina combinabile Hb È un parametro fisiologico importa...

Page 10: ...l polso non sono visualizzati stabilmente 1 Il dito non è inserito abbastanza in profondità 2 Il dito trema o il paziente si sta muovendo 1 Posizionare il dito correttamente e riprovare 2 Lascia che il paziente mantenga la calma Il dispositivo non può essere acceso 1 Batteria scarica o nessuna batteria 2 Le batterie non sono inserite correttamente 3 Il malfunzionamento del dispositivo 1 Sostituire...

Page 11: ...lema Posible razón Solución La SpO2 y la frecuencia del pulso no se muestra normalmente 1 El dedo no estácolocado correctamente 2 El valor de la SpO2 en el paciente es demasiado baja para ser detectado 1 Coloque el dedo correctamente y vuelva a intentarlo 2 Inténtelo de nuevo Vaya al hospital para un diagnó stico si está seguro de que el dispositivo funciona correctamente La SpO2 y la frecuencia d...

Page 12: ...treet London W1G 9QR UK Shenzhen Jumper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China E mail info jumper medical com Tel 86 755 26692192 26696279 Web www jumper medical com ...

Reviews: