Hangzhou Junce Instruments Co., Ltd.
17
value. Press button A, the cursor appears at the current setting position,
rotate the knob to change the value, press the left and right buttons to change
the step value. Press OK to turn the output on or off.
The voltage and current setting display interfaces are as follows:
The output status interface is as follows
:
2.Function Setting
Press the SET button on the main interface, the cursor will appear at the
position of the function option box 01, press the left or right button or turn the
knob to select load, save, clear, address, channel, quick response, automatic
output, upper limit setting, lower limit setting, limit reset, brightness, auto
power off, language, factory reset, about and other 15 function settings, press
the OK button to enter function settings, press the left or right button or turn
the knob to change the parameters, press the OK button to finish the setting.
An example of the function setting display interface is as follows:
Boot default connection status
Voltage setting status
Current setting status
5V, 1A constant voltage
output, when the load current
is less than the set current, the
output is a constant voltage
output, and the CV light is on.
When the load current
reaches the set current, the
output is the set current,
constant current output,
and the CC light is on.